Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 88: (1 year) CAL VER /5700,5725 DATE: 10-Sep-98 AUTHOR: Fluke Corporation REVISION: $Revision: 2.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 57 NUMBER OF LINES: 241 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A ============================================================================= # # Source: # Fluke 88 Service Manual (PN 880906 Dec. 1990, Rev. 2, 9/94). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # Sub Fluke 86,88 (1 year) Frequency CAL VER /3325 # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P J F W 1.002 ASK+ K 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP No warm-up time requirement. 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 90%. 1.006 ASK- U 1.007 HEAD {BASIC METER OPERATIONS} 1.008 DISP Turn the Meter rotary switch to Ohms. 1.009 DISP Connect a test lead from V Ohm Diode input to the 1.009 DISP mA input. (If you are using a test probe, touch 1.009 DISP the half of the input contact nearest the LCD.) 1.010 ACC 1.000kZ TOL 1.011 MEMI Enter UUT reading in kilohms: 1.012 MEME 1.013 MEMC kZ 0.005U 2.001 DISP Remove the test lead from the UUT. 2.002 HEAD {15A FUSE TEST} 2.003 DISP Insert the plug end of the test lead into the A input. 2.004 EVAL Is the UUT beeping repeatedly? 3.001 HEAD {1A FUSE TEST} 3.002 DISP Insert the plug end of the test lead into the mA input. 3.003 EVAL Is the UUT beeping repeatedly? 4.001 DISP Remove the test lead. 4.002 HEAD {DISPLAY TEST} 4.003 DISP With the Meter rotary switch in the OFF position, press 4.003 DISP and hold down any button and turn the rotary switch 4.003 DISP to any function setting. 4.004 EVAL Did the UUT display test pass? 5.001 HEAD {ROTARY SWITCH TESTS} 5.002 DISP Press and hold down RPM button for 2 seconds while 5.002 DISP rotating the UUT function switch from OFF to {VAC}. 5.003 MATH MEM1 = 0 5.004 MEMI Enter UUT reading: 5.005 MEME 5.006 MEMC display 12U 6.001 DISP Rotate the UUT function switch to {VDC}. 6.002 MATH MEM1 = -32 6.003 MEMI Enter UUT reading: 6.004 MEME 6.005 MEMC display 12U 7.001 DISP Rotate the UUT function switch to {mVDC}. 7.002 MATH MEM1 = -64 7.003 MEMI Enter UUT reading: 7.004 MEME 7.005 MEMC display 12U 8.001 DISP Rotate the UUT function switch to {Ohms}. 8.002 MATH MEM1 = -96 8.003 MEMI Enter UUT reading: 8.004 MEME 8.005 MEMC display 12U 9.001 DISP Rotate the UUT function switch to {Diode}. 9.002 MATH MEM1 = -128 9.003 MEMI Enter UUT reading: 9.004 MEME 9.005 MEMC display 12U 10.001 DISP Rotate the UUT function switch to {mA/A DC}. 10.002 MATH MEM1 = -160 10.003 MEMI Enter UUT reading: 10.004 MEME 10.005 MEMC display 12U 11.001 DISP Rotate the UUT function switch to {mA/A AC}. 11.002 MATH MEM1 = -192 11.003 MEMI Enter UUT reading: 11.004 MEME 11.005 MEMC display 12U 12.001 ASK+ U 12.002 HEAD {AC VOLTAGE TESTS} 12.003 DISP Select Autorange VAC by pressing and holding the RANGE 12.003 DISP button for 2 seconds while rotating the UUT function 12.003 DISP switch from OFF to the VAC setting. 12.004 DISP Connect the 5700A and the UUT as follows. 12.004 DISP [32] 5700A OUTPUT HI to UUT V Ohm Diode 12.004 DISP [32] 5700A OUTPUT LO to UUT COM 12.004 DISP [32] 5700A SENSE HI to UUT V Ohm Diode 12.004 DISP [32] 5700A SENSE LO to UUT COM 12.005 HEAD AC VOLTAGE TESTS: {400mV Range} 12.006 5700 400 350.0mV 6.3U 60H 2W 13.001 5700 400 350.0mV 6.3U 1kH 2W 14.001 5700 400 350.0mV 7.7U 5kH 2W 15.001 HEAD AC VOLTAGE TESTS: {4V Range} 15.002 5700 4 3.500V 0.058U 60H 2W 16.001 5700 4 3.500V 0.058U 1kH 2W 17.001 5700 4 3.500V 0.072U 5kH 2W 18.001 HEAD AC VOLTAGE TESTS: {40V Range} 18.002 5700 40 35.00V 0.58U 60H 2W 19.001 5700 40 35.00V 0.58U 1kH 2W 20.001 5700 40 35.00V 0.72U 5kH 2W 21.001 HEAD AC VOLTAGE TESTS: {400V Range} 21.002 5700 400 350.0V 5.8U 60H B1 2W 22.001 5700 400 350.0V 5.8U 1kH B1 2W 23.001 5700 400 350.0V 7.2U 5kH B1 2W 24.001 HEAD AC VOLTAGE TESTS: {1000V Range} 24.002 5700 1000 900V 28U 60H B1 2W 25.001 5700 1000 900V 28U 5kH B1 2W # Better TURs for the frequency tests can be achieved using a HP 3325B # function generator. If you wish to use the function generator in lieu # of the 5700A delete the 5700A based frequency tests and uncomment the # CALL statement at the bottom of this procedure. 26.001 HEAD {FREQUENCY TESTS} 26.002 DISP Press the RANGE button to lock the 400mV Range. 26.002 DISP Press RPM three times to choose the Frequency mode. 26.003 HEAD FREQUENCY TESTS: {19.999kHz Range} 26.004 5700 20 19.000kH 0.003U 150mV 2W #! WARNING: Test Tol 3, Sys Tol 1.9, TUR 1.5789 (< 4.00). 27.001 HEAD FREQUENCY TESTS: {199.99kHz Range} 27.002 5700 200 190.00kH 0.03U 150mV 2W #! WARNING: Test Tol 30, Sys Tol 19, TUR 1.5789 (< 4.00). 28.001 HEAD {FREQUENCY SENSITIVITY AND TRIGGER LEVEL} 28.002 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {4V AC Range} 28.003 DISP Press the RANGE button to lock the 4V Range. 28.004 5700 1000.0H 0.2U 300mV 2W #! WARNING: Test Tol 0.2, Sys Tol 0.1, TUR 2.0000 (< 4.00). 29.001 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {4V DC Range} 29.002 DISP Rotate the UUT function switch to VDC. 29.002 DISP Press RANGE to select the 4V Range. 29.002 DISP Press RPM three times to choose the Frequency mode. 29.003 5700 1000.0H 0.2U 1.7V 2W #! WARNING: Test Tol 0.2, Sys Tol 0.1, TUR 2.0000 (< 4.00). 30.001 5700 1000.0H 1.0V S 2W 30.002 EVAL Is the UUT {displaying all zeros}? 31.001 HEAD FREQUENCY SENSITIVITY AND TRIGGER LEVEL: {40V DC Range} 31.002 DISP Press the RANGE button to lock the 40V Range. 31.003 5700 1000.0H 0.2U 6.0V 2W #! WARNING: Test Tol 0.2, Sys Tol 0.1, TUR 2.0000 (< 4.00). 32.001 5700 1000.0H 2.0V S 2W 32.002 EVAL Is the UUT {displaying all zeros}? 33.001 HEAD {DC VOLTAGE TESTS} 33.002 DISP Press the RPM button to select VDC. 33.002 DISP Press the RANGE for 2s to select autorange. 33.003 HEAD DC VOLTAGE TESTS: {4V Range} 33.004 5700 4 3.500V 0.007U 2W 34.001 HEAD DC VOLTAGE TESTS: {40V Range} 34.002 5700 40 35.00V 0.07U 2W 35.001 5700 40 -35.00V 0.07U 2W 36.001 HEAD DC VOLTAGE TESTS: {400V Range} 36.002 5700 400 350.0V 0.7U 2W 37.001 HEAD DC VOLTAGE TESTS: {1000V Range} 37.002 5700 1000 1000V 5U 2W 38.001 HEAD {DC MILLIVOLT TEST} 38.002 DISP Rotate the UUT function switch to mVDC. 38.003 HEAD DC MILLIVOLT TEST: {400mV Range} 38.004 5700 400 350.0mV 0.7U 2W 39.001 HEAD {RESISTANCE TESTS} 39.002 DISP Rotate the UUT function switch to Ohms. 39.003 HEAD RESISTANCE TESTS: {400 Ohm Range} 39.004 5700 400 190.0Z 0.5U CW 40.001 HEAD RESISTANCE TESTS: {40 kOhm Range} 40.002 5700 40 19.00kZ 0.05U CW 41.001 HEAD RESISTANCE TESTS: {4 MOhm Range} 41.002 5700 4 1.900MZ 0.005U 2W 42.001 HEAD RESISTANCE TESTS: {40 MOhm Range} 42.002 5700 40 19.00MZ 0.22U 2W 43.001 HEAD RESISTANCE TESTS: {400 MOhm Range} 43.002 DISP Press the Range button twice for 400 MOhm range. 43.003 5700 400 100MZ 3U 2W 44.001 HEAD {DIODE TEST} 44.002 DISP Rotate the UUT function switch to Diode. 44.003 5700 3.000V 0.061U 2W 45.001 HEAD {DC MILLIAMP TESTS} 45.002 DISP Rotate the UUT function switch to mA/A DC. 45.003 DISP Connect the 5700A and the UUT as follows. 45.003 DISP [32] 5700A OUTPUT HI to UUT mA 45.003 DISP [32] 5700A OUTPUT LO to UUT COM 45.004 HEAD DC MILLIAMP TESTS: {40mA Range} 45.005 5700 40 35.00mA 0.30U 2W 46.001 HEAD DC MILLIAMP TESTS: {400mA Range} 46.002 5700 400 350.0mA 3.0U 2W 47.001 HEAD {AC MILLIAMP TESTS} 47.002 DISP Rotate the UUT function switch to mA/A AC. 47.003 HEAD AC MILLIAMP TESTS: {40mA Range} 47.004 5700 40 35.00mA 1.15U 60H 2W 48.001 5700 400 35.00mA 1.15U 1kH 2W 49.001 HEAD AC MILLIAMP TESTS: {400mA Range} 49.002 5700 400 350.0mA 5.8U 60H 2W 50.001 5700 400 350.0mA 5.8U 1kH 2W 51.001 DISP Disconnect the 5700A from the UUT. 51.002 HEAD {AC AMP TESTS} 51.003 DISP Connect the 5725A and UUT as follows: 51.003 DISP [32] 5725A CURRENT OUTPUT HI to UUT A 51.003 DISP [32] 5725A CURRENT OUTPUT LO to UUT COM 51.004 HEAD AC AMP TESTS: {4000mA Range} 51.005 5700 4000 3500mA 58U 60H B1 2W 52.001 5700 4000 3500mA 58U 1kH B1 2W 53.001 HEAD AC AMP TESTS: {10A Range} 53.002 5700 10 10.00A 0.20U 60H B1 2W 54.001 5700 10 10.00A 0.20U 1kH B1 2W 55.001 HEAD {DC AMP TESTS} 55.002 DISP Rotate the UUT function switch to mA/A DC. 55.003 HEAD DC AMP TESTS: {4000mA Range} 55.004 5700 4000 3500mA 30U B1 2W 56.001 HEAD DC AMP TESTS: {10A Range} 56.002 5700 10 10.00A 0.15U B1 2W #57.001 CALL Sub Fluke 86,88 (1 year) Frequency CAL VER /3325 57.001 END #! T.U.R.s less than 4.00: 5 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 8 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.